Contrary electrode positions and its effect on the accuracy of apex locator


Author Details : Logani A, Singh A, Chawla A, Kumar V, Shah N

Volume : 3, Issue : 1, Year : 2015

Article Page : 40-42


Suggest article by email

Get Permission

Abstract

Aim: Clinically evaluate the effect of varying the conditions and the position of the contrary electrode on the accuracy of apex locators to determine the working length.
Methodology: Thirty patients requiring endodontic intervention of as many maxillary anterior teeth were randomly selected. Root ZX, a 3rd generation apex locator was used to determine the electronic working length (EWL). Three positions of the contrary electrode were evaluated Group I: held at the corner of the patient’s mouth in contact with oral mucous membrane, Group II: held in the patients’ dry hand, Group III: held in patients moist hands. Three consecutive readings were taken at each position and the mean calculated.
Results: The readings obtained for group II & III were analyzed against group I using Altman Criterion for Agreement Analysis. A mean difference of 0.73 and 6.52mm was observed between group I & III and between groups I & II respectively.
Conclusion: Moist hands can be an alternate placement position for the contrary electrode during EWL determination.


How to cite : Logani A, Singh A, Chawla A, Kumar V, Shah N, Contrary electrode positions and its effect on the accuracy of apex locator. J Dent Spec 2015;3(1):40-42


This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.







View Article

PDF File  


Copyright permission

Get article permission for commercial use

Downlaod

PDF File    






Article Access statistics

Viewed: 1543

PDF Downloaded: 394